Particle Counter Type | Liquid |
Brand | yateks |
Model Name/Number | YPA-FJ4 |
Usage/Application | Industrial |
Type Of Data Logger | Digital Data Logger |
Test parameters | Viscosity DensityOil Quality(Dielectric constant)Temperature |
Oil sample volume | 10ml |
Built-in printer | Convenient print to test report |
USB port | Export test records, software update |
RS232 port | Modbus RTU communication promotol |
Storage | 32G memory, can store 100K records |
Light source Semiconducting laser narrow light detector (more accurate, steady and rapid)Particle size range 1μm~100μmSensitivity 1μm (ISO4402) or 4μm(C) (ISO11171)Off-line viscosity detection ≤160cStTest accuracy ±1 pollution level.Detection channelComply to ISO 4406, NAS1638 standard;Size range: 1um–100um;8 channels: NAS1638 standard: 5um/15um/25um/50um/100um;ISO4406 standard: 4um/6um/14umOil temperature: 0-100℃; Accuracy:±0.5℃;Humidity: 0-100%; Accuracy:±3%;Sampling volume 10mlPower supply 220V 50Hz (Rechargeable battery for 4-6 hours working)Testing data output Equip with a printer and support USB download historical data.Dimension 273*200*134mmNet weight 6kgOil temperature 0℃~80℃